Figure 86: The result of DICOM Measurement
Using this function:
If you want to enable this function, the following actions are required:
(1) Unzip the GammaCompMD QA package into a local or network hard disk folder, from
where the installation will be started.
(2) Open the StandaloneCalibration.ini file in the same folder as the installer setup.exe with
a text editor.
(3) Change the value of the line StandAloneCalibration=0 into StandAloneCalibration=1 and
save the file.
(4) Install GammaCompMD QA.
This function can be enabled or disabled only at the time of installation.
If you want to disable this function again, the following actions are required:
(1) If required, do a Backup of all data before you uninstall GammaCompMD QA.
(2) Uninstall GammaCompMD QA.
(3) Open the StandaloneCalibration.ini file in the same folder as the installer setup.exe with
a text editor.
(4) Change the value of the line StandAloneCalibration=1 into StandAloneCalibration=0 and
save the file.
(5) Re-install GammaCompMD QA again. If required, restore the data after re-installation.