GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 180
Regular re-adjustment of integrated front sensors or retractable sensors
It is a common physical phenomenon that all light sensors lose their sensitivity over time
when a light source emitting some ultraviolet light is measured permanently. The backlights
of LCD Displays contain some small part of ultraviolet light, causing the plastic cover material
of such a sensor to deteriorate. The sensitivity of the sensor will be reduced over time. This
results in reduced measurement accuracy when performing i.e. a DICOM conformance test.
A Reference Calibration provides a re-adjustment of the front sensor and improves the
measurement accuracy again. In case of displays with integrated front sensors, the new
reference values are stored in the display.
Recommended time interval for Reference Calibration
At time of installation and first calibration
Then regularly every 24 months in case of a high amount of active usage
hours of the display, shorter time intervals are recommended
MD215MG (5MP) regularly every 12 months in case of high active
usage hours of the display, shorter time intervals are recommended
NOTE: In case of models MD212G3, MD215MG, MD211G5 and MD302C6, a Reference
Calibration of the front sensor is performed automatically during calibration with an external
sensor. Therefore the Reference Calibration button, as described in 7.2.6 Sensor Setup(page
77) is disabled for these models.