GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 108
In case of confusion, the Automatic Configuration feature may be used to restore the
original settings and then to start the tests again with a defined display arrangement.
NOTE: The Automatic Configuration function cannot detect whether one of the connected
displays is the Navigation or RIS monitor of a PACS system. Any display which cannot be
detected as a NECDisplay will automatically be shown as a NAVDisplay.
However, it may be helpful to manually change the Display Interface Mode for a third party
display to StdDisplay within this Initialize Display Configuration process, in order enable
functions like DICOM conformance tests or QA tests.
7.7.1.2. Reinitialize System Configuration
When Reinitialize System Configuration is clicked in Administrator, the Information dialog
box, as shown in Figure 91, is displayed. (Local) System Administrator access rights are
required to initialize the system configuration.
Figure 91: Information dialog box
Clicking the OK button will either display a confirmation dialog box reminding about system
administrator access rights, or a dialog box for raising the access rights level, depending on
the operating system. Follow the instructions to complete this step. Then the Initialize
System Configuration dialog box, as shown in Figure 92, will be displayed.