GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 60
7.2.4. Schedule Setup
Click on Schedule Setup in Calibration to display the Schedule Setup dialog box,
see Figure 39.
NOTE: Scheduled action can use not only Display Sensor but also External Sensor.
Display Sensor will be used preferentially regardless of the Preferred Sensor Selection
Settings. See 7.2.6 Sensor Setup (page 77)
If the display has no Display Sensor, External Sensor will be used.
The following External Sensors are available
MDSVSENSOR3 by NEC (USB)
i1 Display version 2 by X-Rite (GretagMacbeth) (USB)
i1Display Pro by X-Rite (USB)
Spyder3 by Colorvision (USB)
Unfors Luxi (RS-232C) - Color measurement not supported
IBA LXcan(USB) - Color measurement not supported
IBA LXchroma(USB)
Note for the displays which have no Display Sensor:
When you use the External Sensor, put it on the center of the screen beforehand.
When the External Sensor is in the ambient light measurement mode, the scheduled
action will fail.
Only one External Sensor is available.
The following cases are not executed.
External Sensor is not put on the screen beforehand.
External Sensor is put on the screen of the display excluding for schedule action.
External Sensor is not connected.