GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 81
A LCD display is fragile and may be damaged if the external sensor
is forcefully pressed against it. Never attach a sensor with suction
cups to the display. Always use the supplied cable and suspend the
sensor with a weight so that the sensor is stationary in front of the
display.
If the external sensor is separated from the screen during
calibrations or tests (i.e. by falling off), the process may fail. Be
sure to keep the external sensor in place, until the process is
completed. If the sensor has been displaced, position the sensor
on the screen again and best restart the process.
Figure 57: Sensor Contact Position Guide
Black Level setup Several supported external sensor models require the adjustment
of the Black Level (Dark Current) for improved accuracy. This Black
Level measurement will be performed before any calibrations and
tests. Please follow the guidance on the screen. In most cases, the
sensor needs to be placed on a dark surface for this setup.
Figure 58: X-Rite Chroma 5 during black level adjustment (Example)