GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 73
Use/Don’t use Ambient Light Compensation
Select between Don’t use Ambient Light Compensation or Use
Ambient Light Compensation. Ambient Light Compensation is
active during Calibrations, Conformance Tests, White/Black
Luminance Measurements and Uniformity Tests.
Illumination Shows the value of illumination (ambient light, unit: lux) as
measured by the display’s ambient light sensor at time of
installation. The displayed value (1 to 1000 lux) may be modified to
a value measured with an external Illuminance meter (Lux meter).
Reflected luminance The Illumination value is used to calculate and display the
Reflected luminance (Unit: cd/m
2
). If you want to edit the
Reflected luminance, click on the Manually adjust the
Illumination value to modify the Reflected Luminance value
button. The Reflected luminance Setup dialog box will be shown
similar to Figure 52.
Manually adjust the Illumination value to modify the Reflected Luminance value
The Illumination value and the calculated Reflected Luminance
value are displayed in the Reflected Luminance Setup dialog box,
as shown in Figure 52. In order to update the Reflected Luminance
value, click this button to modify this value.
Figure 52: Reflected Luminance Setup dialog box
Select the check box to modify the Reflected luminance value
manually (Range: 0.01 to 9.99 cd/m
2
). Click on the OK button to
calculate the Diffuse Reflection value and transfer this value back
into the display. No modification will happen if the check box
remains unchecked. The Illumination value is not updated.