GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 139
NOTE:
・The luminance for the uniformity test reports refers to the output luminance for the target
curve designated in the Grayscale Function tab within 7.2.5 Calibration Setup (page 64).
・An external sensor is required for uniformity testing. (Even if a display sensor was selected,
the external sensor is automatically used.) If an external sensor is not connected or other
application is using external sensor, the Figure 126 is displayed.
Figure 126:Warning dialog box
7.7.3.4. Display Matching
When you click on Display Matching in Administrator, the Display Matching dialog box
Figure 127 is displayed. Display matching is a function to adjust the luminance, color
temperature and gamma correction curve from one display to one or more displays. The goal
is that a target display has the same luminance, color temperature and gamma correction
curve as the source display. An external sensor is required to perform display matching.
Please refer to 7.2.6 Sensor Setup (page 77) regarding correct sensor settings.
Figure 127: Display Matching dialog box